Principle
Fluorescence radiation of the elements of a sample can cause
fluorescence radiation inside the detector and its housing if the
energy is sufficiently high. As a result, the spectrum may include
lines that are not caused by the sample. For the detection of
potential additional lines, the detector is subjected to
monochromatic X-radiation with the aid of a monocrystal. For
comparison, the fluorescence spectra of pure metal samples are
measured.
Tasks
Calibrate the semiconductor energy detector with the aid of the
characteristic fluorescence radiation of the calibration
sample.
Irradiate the X-ray energy detector with monoenergetic X-rays
that are produced by the Bragg reflection on an LiF monocrystal.
Measure the resulting fluorescence spectrum.
Determine of the energy of the spectrum lines.
Assign the lines to elements by comparing the measured values
with table values.
Comparative measurement and evaluation of the fluorescence
spectra of pure metal samples.
What you can learn about
Bremsstrahlung
Characteristic X-radiation
Fluorescence radiation
Fluorescent yield
Interference of X-rays
Crystal structures
Bragg’s law
Compton scattering
Escape peaks
Semiconductor energy detectors
Multichannel analysers