ÜRÜN BİLGİLERİ
Principle
Various ore samples are subjected to polychromatic X-rays.
The energy of the resulting fluorescence radiation is analysed with
the aid of a semiconductor detector and a multichannel analyser.
The energy of the corresponding characteristic X-ray fluorescence
lines is determined. The elements of the samples are identified by
comparing the line energies with the corresponding table
values.
Tasks
Calibrate the semiconductor energy detector with the aid of the
characteristic radiation of the tungsten X-ray tube.
Record the fluorescence spectra that are produced by the
samples.
Determine the energy values of the corresponding fluorescence
lines and compare the experimental energy values with the
corresponding table values in order to identify the ore
samples.
What you can learn about
Bremsstrahlung
characteristic X-radiation
energy levels
fluorescent yield
semiconductor energy detectors
multichannel analysers